Return Home

Sutter Instruments P-2000 Cook Book

This is a compilation of Sutter Instruments P-2000 Pipette Puller parameters that I find in the literature, create myself, or have received from someone else. If you would like to contribute, please email me.

Single-Barrel Capillaries

Diameter Material Brand Dimensions Line Number HEAT FIL VEL DEL PUL Reference
50 nm Quartz Sutter Instruments 1.0 OD X 0.5 ID X 100 L mm 1 750 4 30 150 80 Ref
2 650 3 40 135 150
800 nm Borosilicate Harvard Apparatus 1.2 OD X 0.69 ID X 100 L mm 1 800 4 40 130 30 Ref
2 750 3 20 130 100
400 nm Borosilicate Harvard Apparatus 1.2 OD X 0.69 ID X 100 L mm 1 350 3 40 220 0 Ref
2 350 3 40 180 120
150 nm Borosilicate Harvard Apparatus 1.2 OD X 0.69 ID X 100 L mm 1 350 3 30 220 0 Ref
2 350 3 40 180 120
200 nm Borosilicate Harvard Apparatus 1.2 OD X 0.69 ID X 100 L mm 1 330 3 30 220 0 Ref
2 330 3 40 180 120
150 nm Borosilicate Harvard Apparatus 1.2 OD X 0.69 ID X 100 L mm 1 350 3 30 220 0 Ref
2 350 3 40 180 120
200 nm Borosilicate Harvard Apparatus 1.2 OD X 0.69 ID X 100 L mm 1 330 3 30 220 0 Ref
2 330 3 40 180 120
60 nm Borosilicate Harvard Apparatus 1.2 OD X 0.69 ID X 100 L mm 1 350 3 30 220 0 Ref
2 350 3 40 180 120